Dissolved Oxygen Transmitter WE-T123 Series

     Product Features

    Split design

    • The transmitter body adopts LoRa type to support 410MHz~490MHz band range and SMA antenna interface; 4G type supports full network communication;
    • The transmitter communicates with the edge computing gateway of Shidian Technology through the WES-WIO wireless terminal communication protocol to realize the real-time reporting of the dissolved oxygen content and status detection data in the field liquid;
    • External dissolved oxygen probe with optional mounting;

     

    dissolved oxygenProbe Specifications

    • 5m cable length, support cable length customization;
    • Adopts the fluorescence method of measurement principle, based on the physics of specific substances on the excitation of fluorescence burst principle design;
    • Easy to install, 3/4 NPT for countersunk installation or mounting in pipes and tanks;

     

    Robust industrial-grade design

    • IP67 protection rating for the main unit and IP68 protection rating for the probe;
    • The mainframe supports -40 to 85° storage temperature, -20 to 70° working ambient temperature, and 5 to 95% relative humidity;
    • Probe operating conditions 0 ~ 50 ℃, ≤ 0.2MPa, storage temperature -5 ~ 65 ℃

     

    Thorough installation and after-sales maintenance

    • The mainframe supports mounting methods such as rail, wall mounting, snap ring, hoop and so on;
    • The P0 model supports a 19,000mAh high-capacity lithium battery with a typical endurance of 5 years for a 5-minute reporting cycle and up to 15 years for a 1-hour reporting cycle;
    • Supports rich fault detection functions, including dissolved oxygen threshold exceeding the limit, abnormal changes in dissolved oxygen, low battery and other fault detection, and notifies after-sales maintenance personnel by cell phone text messages and emails through the equipment management platform of Shidian Technology;
    • Support high definition LCD display, with LED indicator and keypad, you can view the dissolved oxygen value and alarm information in real time on site;

    ¥2,248¥2,373

    WE-T123 is a transmitter used to monitor the dissolved oxygen content in water, and its transmission method covers LoRa, 4G, NB-IoT and other ways, while the power supply method is divided into two types: battery-powered and 24V-powered. The dissolved oxygen transmitter adopts a highly stable digital sensor, which is capable of real-time monitoring and calculating the liquid on site. Through LoRa, 4G, NB and other communication technologies, it realizes long-distance data transmission and combines with the industrial edge computing gateway products of WorldTech to upload the data to the equipment management platform of WorldTech, so as to complete the real-time collection, analysis and calculation, storage and display of the dissolved oxygen in the on-site liquid.

    Transmitter body and probe split design, for the site of the measured liquid, as well as installation methods and other different needs, to provide a variety of installation methods. Widely used in wastewater treatment, clean water monitoring, fermentation tanks, ultrapure water, brewing and other industrial sites.

    transmission method

    , , ,

    Power supply method

    ,

    specification

    functional category Functional Description note
    communication interface

     

    LORA communication port

     

    Antenna connector type: SMA female

    Support 410.125~490.125MHz band

    22dbm maximum transmit power.

    Maximum -147dbm reception sensitivity, open line-of-sight 5KM signal coverage capability

    4G

     

     

    MQTT, HTTP, WES cloud service protocols

    LTE FDD B1/3/5/8

    LTE TDD B34/38/39/40/41 (full band)

    Rate (Mbps) LTE FDD: 10.3(DL)/5.1(UL)

    LTE TDD: 9.1 (DL)/3.1 (UL)

    NB

     

    LTE Cat NB2: B3/B5/B8

    Rate (kbps): Single-Tone: 25.5 (DL)/16.7 (UL)

    Multi-Tpne: 127(DL)/158.5(UL)

    UI 1* Key input Refer to UI specification definition, multi-function buttons
    2* LED Indicators System, LoRa Communication Indicators
    screen (TV, computer or movie) 128*64 dot matrix monochrome LCD

    Display of sensor values and status information

    Setting up interactions

    electricity supply Lithium battery power supply 19000mAh, Li-Ion Battery
    24VDC power supply Wide voltage input, 9 to 30V
    matrix Mainframe working environment Working temperature: -20~70℃
    Operating relative humidity: 5~95%
    Host Storage Environment Storage temperature: -40~85℃
    Storage relative humidity: 5 ~95%
    Probe working conditions 0~50℃, ≤0.2MPa
    Probe storage temperature -5 to 65°C
    Protection level Waterproof and dustproof mainframe IP67
    ESD 4KV contact, 8KV non-contact
    Probe waterproof and dustproof IP68
    physical property Main unit exterior dimensions 78(W)*121(H)*51.4(D)mm
    Main unit mounting method Wall-mounted, pole, hoop, rail mounted
    monitors Measurement type dissolved oxygen
    Measurement range Dissolved oxygen: 0~20.00 mg/L (0~200% saturation, 25℃)
    Detection accuracy ±2%
    minimum resolution 0.01mg/L
    Fluorescent Film Head Life 1 year (under normal use)
    Probe mounting method Countersunk mount, 3/4 NPT

    Related drawings

    Dimensions|Mounting

    network topology

    Models and accessories

    Resource support

    WE-T123 Dissolved Oxygen Transmitter Product Specification